MOCON Oxygen Transmission Analyzer
OX-Tran 2/12 OTR Analyzer (R model)
Applications:
Packaging Films
- Monolayer
- Multi-layer
- Coated Films
- Coated Paper/paperboard
Specialty Films
- Gel/lotions
Packages
- Common food packages
- Medical packages
- Bottles/Closures
Test Standard Compliance
ASTM F2622
ISO CD 15105-2
JIS K-7126-B
DIN 53380-3
Technical Specifications:
Capabilities – Model Specific
Units | R Model | ||
Test Ranges (50 cm^2 @100%O2) | cc/(m2 × day) | 0.05 to 28,800 | |
cc/(100 in2 × day) | 0.003 to 1858 | ||
cc/(pkg × day) | 0.00025 to 144 | ||
Test Ranges (5.6 cm^2) @100%O2 | cc/(m2 × day) | 0.44 to 255,300 | |
cc/(100 in2 × day) | 0.027 to 16,470 | ||
cc/(pkg × day) | 0.0022 to 1277 | ||
Resolution | cc/(m2 × day) | 0.05 | |
Repeatability | cc/(m2 × day) | 0.05 or 1.5%* | |
Test Temperature Range | 10° to 40°C ± 0.2°C | ||
Controlled RH Testing Range | 0 to 90% ± 3% |
*whichever is greater
Capabilities – All Models
Ox-Tran 2/12 Series | Units |
Film Size | 4” x 4” Std, 2”x2” optional |
Film Thickness | Up to 120 Mil |
Film Test Area | 50 cm2 cartridge std, 5.6 cm2 cartridge optional |